next up previous contents index
Next: Conventional Theories Up: Non-hydrodynamic effects in Previous: Non-hydrodynamic effects in

Steady State Townsend Experiments

In these experiments, the source of charged particles is constant in time [Huxley and Crompton [1974]] . The usual transport coefficients measured by   this technique are the first Townsend ionization coefficient, and diffusion perpendicular to the field (lateral diffusion).   The ionization coefficient is measured in a parallel plane   apparatus where the variation of the source in the radial direction can be ignored (figure gif).

The current obtained at the collector is measured as a function of the distance d between the electrodes. If charged particles are ionizing the background gas, then the density of charged particles in the drift region is found to vary exponentially   as , where is known as the first Townsend ionization coefficient.  

  
Figure: Parallel Plane Steady State Townsend Experiment

Lateral diffusion is measured by means of a split collector (fig gif). In this case, the source has small diameter compared with the dimensions of the apparatus (idealized as a point source). The diffusion perpendicular to the field is extracted from the ratio of currents measured on section A and B. Before 1967, it was assumed that diffusion had   the same value in all directions. That this is not the case was shown by Wagner, Davis and Hurst [1967] who from time of flight experiments obtained coefficients of diffusion parallel to the field direction that were different to those obtained by the Townsend-Huxley method.  

  
Figure: Townsend-Huxley Diffusion Experiment. Ratio of currents recieved by sections A and B are measured.



Russell Standish
Thu May 18 11:43:52 EST 1995